CVE-2017-12718
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.…
- 8.1HighCVSS base score
- 12.8%Exploitation in 30 daysEPSS · top 3.7% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- 1Public exploitsMetasploit, Nuclei and Exploit-DB
Description
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
Public exploit
An exploit or proof of concept is public: anyone can try it against an unpatched system. The public collections list an Exploit-DB entry for it. A reference of the record is tagged as an exploit.
Show the exploits2
Exploit-DB1 entry
- Smiths Medical Medfusion 4000 - 'DHCP' Denial of Service (Exploit-DB entry) (opens in a new tab)
EDB-ID 43776dos, hardwarepublished 18 Jan 2018Not verified
References tagged exploit1 reference
- 43776 (opens in a new tab)exploitx_refsource_EXPLOIT-DBx_transferred
Modules, templates and exploits as listed by the Metasploit Framework (BSD-3-Clause), ProjectDiscovery's Nuclei templates (MIT) and the Exploit Database (GPLv2), read once a day. Where the data comes from
Timeline
ID reserved
By icscert
- 6 months later
· 6 months later
Published
- 6 years later
· 6 years later
Record updated
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| NVDNVD · Modified | 3.0 | High8.1 | CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 12.8% · top 3.7% of all CVEs (10 Nov 2026)
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.1.
Affected products
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
- affected Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
References
4
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A (opens in a new tab)x_refsource_MISCx_transferred
- 101252 (opens in a new tab)vdb-entryx_refsource_BIDx_transferred
- 100665 (opens in a new tab)vdb-entryx_refsource_BIDx_transferred
- 43776 (opens in a new tab)exploitx_refsource_EXPLOIT-DBx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2017-4257 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket