Skip to content
CVE Ledger
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • MCP serverPage
  • Data sourcesPage
  • Data statusPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
  • Support the projectPage
Back

CVE-2017-12726

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and…

Description

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

Timeline

  1. ID reserved

    By icscert

  2. · 6 months later

    Published

  3. · 6 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2017-12726, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.0High7.3CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
EPSS (exploitation in 30 days)
1.1% · top 35% of all CVEs (10 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.3.

Affected products

  • Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    • affected Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

References

2

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2017-12726 badge
[![CVE-2017-12726](https://cveledger.app/badge/CVE-2017-12726.svg)](https://cveledger.app/cve/CVE-2017-12726)

Record last updated .