CVE-2018-14781

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Description

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Timeline

  1. ID reserved

    By icscert

  2. · 12 days later

    Published

  3. · 7 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2018-14781, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1Medium5.3CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
NVDNVD · Modified3.1Medium5.3CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
EPSS (exploitation in 30 days)
0.71% · top 46% of all CVEs (11 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 5.3.

Affected products

What to do

Workaround: The remote option is turned off in the pump by default.   Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic. Medtronic has released additional patient focused information https://www.medtronic.com/security . Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.