Skip to content
CVE Ledger
  • Newest CVEsPage
  • Recently added to KEVPage
  • Rising exploitation (EPSS)Page
  • Weekly digestPage
  • StatisticsPage
  • WatchlistPage
  • API documentationPage
  • MCP serverPage
  • Data sourcesPage
  • Data statusPage
  • ChangelogPage
  • AboutPage
  • How it's builtPage
  • Support the projectPage
Back

CVE-2018-8870

Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

Timeline

  1. ID reserved

    By icscert

  2. · 3 months later

    Published

  3. · 7 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2018-8870, by who scored it
Scored byCVSSScoreVector
icscertCNA (assigner)3.1Medium6.4CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
NVDNVD · Modified3.1Medium6.4CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS (exploitation in 30 days)
0.36% · top 69% of all CVEs (10 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.4.

Affected products

What to do

Solution: Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security

Workaround: Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities.   * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. * Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security

References

2

Sources

The same CVE in the primary databases and aggregators, to check against.

Badge for a README or a ticket

CVE-2018-8870 badge
[![CVE-2018-8870](https://cveledger.app/badge/CVE-2018-8870.svg)](https://cveledger.app/cve/CVE-2018-8870)

Record last updated .