CVE-2019-18256

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An…

Description

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Timeline

  1. ID reserved

    By icscert

  2. · 8 months later

    Published

  3. · 4 years later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2019-18256, by who scored it
Scored byCVSSScoreVector
NVDNVD · Modified3.1Medium4.6CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS (exploitation in 30 days)
0.36% · top 70% of all CVEs (11 Nov 2026)
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 4.6.

Affected products

  • BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM

    • affected CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20