CVE-2021-42744Philips MRI 1.5T and 3T Information Exposure
- 5.9MediumCVSS base score
- 0.28%Exploitation in 30 daysEPSS · top 80% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
Philips MRI 1.5T and MRI 3T Version 5.3 through 5.8.1 does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
Timeline
ID reserved
By icscert
- 8 days later
· 8 days later
Published
- 4 years later
· 4 years later
Record updated
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| PhilipsCNA (assigner) | 4.0 | Medium5.9 | CVSS:4.0/AV:L/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N |
| NVDNVD · Modified | 4.0 | Medium5.9 | CVSS:4.0/AV:L/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
- EPSS (exploitation in 30 days)
- 0.28% · top 80% of all CVEs (17 Nov 2026)
The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 5.9.
Affected products
What to do
Workaround: Philips released a software upgrade version 5.8.2 to remediate these vulnerabilities and can be referenced by FCO78100619. As an interim mitigation to these vulnerabilities, Philips recommends the following: Users should operate all Philips deployed and supported products within Philips authorized specifications, including physical and logical controls. Only allowed personnel are permitted in the vicinity of the product. Refer to the Philips instructions for use (IFU) available on InCenter https://incenter.medical.philips.com. Users with questions about their specific MRI product should contact a Philips service support team or regional service support. Philips contact information is available at the Philips customer service solutions website http://philips.com/productsecurity or by calling 1-800-722-9377. For more information regarding these vulnerabilities, see the Philips product security advisory website http://philips.com/productsecurity. Users can also visit the Philips product security website for the latest security information for Philips products.
References
References2
- https://us-cert.cisa.gov/ics/advisories/icsma-21-313-01 (opens in a new tab)x_refsource_MISCx_transferred
- https://www.usa.philips.com/healthcare/about/customer-support/product-security (opens in a new tab)x_refsource_MISCx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2021-29703 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket