CVE-2021-43935ICSMA-21-343-01 Hillrom Welch Allyn Cardio Products
- 8.1HighCVSS base score
- 1.1%Exploitation in 30 daysEPSS · top 33% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.
Timeline
ID reserved
By icscert
- 29 days later
· 29 days later
Published
- 3 years later
· 3 years later
Record updated
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| icscertCNA (assigner) | 3.1 | High8.1 | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVDNVD · Modified | 3.1 | Critical9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 1.1% · top 33% of all CVEs (15 Nov 2026)
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.1.
Affected products
Hillrom · Welch Allyn Q-Stress Cardiac Stress Testing System
- affected 6.0.0 – ≤ 6.3.1
Hillrom · Welch Allyn X-Scribe Cardiac Stress Testing System
- affected 5.01 – ≤ 6.3.1
Hillrom · Welch Allyn Diagnostic Cardiology Suite
- affected 2.1.0
Hillrom · Welch Allyn Vision Express
- affected 6.1.0 – ≤ 6.4.0
Hillrom · Welch Allyn H-Scribe Holter Analysis System
- affected 5.01 – ≤ 6.4.0
Hillrom · Welch Allyn R-Scribe Resting ECG System
- affected 5.01 – ≤ 7.0.0
Hillrom · Welch Allyn Connex Cardio
- affected 1.0.0 – ≤ 1.1.1
What to do
Solution: Hillrom recommends users upgrade to the latest product versions when updated products are available. Information on how to update these products to their new versions can be found on the Hillrom disclosure page.
Workaround: -Disable the SSO feature in the respective Modality Manager Configuration settings. Please refer to the instructions for use (IFU) and/or service manual for instructions on how to disable SSO. -Apply proper network and physical security controls. -Apply authentication for server access.
References
References1
- https://www.cisa.gov/uscert/ics/advisories/icsma-21-343-01 (opens in a new tab)x_refsource_MISCx_transferred
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2021-30797 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://cveledger.app/cve/CVE-2021-43935)Record last updated .