CVE-2022-26388Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices

Description

A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior.

Timeline

  1. ID reserved

    By Baxter

  2. · 3 years later

    Published

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2022-26388, by who scored it
Scored byCVSSScoreVector
BaxterCNA (assigner)3.1Medium6.4CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L
NVDNVD · Deferred3.1Medium6.4CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L
EPSS (exploitation in 30 days)
0.27% · top 81% of all CVEs (12 Nov 2026)
CISA SSVC
Exploitation: none · Automatable: no · Technical impact: total
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.4.

Affected products

What to do

Solution: Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ .

Workaround: Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service.