CVE-2024-12248Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor
- 9.3CriticalCVSS base score
- 1.3%Exploitation in 30 daysEPSS · top 29% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.
Timeline
ID reserved
By icscert
- 56 days later
· 56 days later
Published
- 13 days later
· 13 days later
Record updated
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| icscertCNA (assigner) | 4.0 | Critical9.3 | CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N |
| icscertCNA (assigner) | 3.1 | Critical9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVDNVD · Deferred | 4.0 | Critical9.3 | CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
- EPSS (exploitation in 30 days)
- 1.3% · top 29% of all CVEs (19 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: yes · Technical impact: total
The list uses the assigner's score when there is one, then CISA's, then NVD's: Critical 9.3.
Affected products
Contec Health · CMS8000 Patient Monitor
- affected Firmware version smart3250-2.6.27-wlan2.1.7.cramfs
- affected Firmware version CMS7.820.075.08/0.74(0.75)
- affected Firmware version CMS7.820.120.01/0.93(0.95)
What to do
Solution: Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .
References
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2024-50712 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://cveledger.app/cve/CVE-2024-12248)Record last updated .