CVE-2024-1275Vulnerability in Baxter Welch Allyn Connex Spot Monitor

Description

Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.

Timeline

  1. ID reserved

    By Baxter

  2. · 4 months later

    Published

  3. · 2 months later

    Record updated

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2024-1275, by who scored it
Scored byCVSSScoreVector
BaxterCNA (assigner)4.0Critical9.1CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N
NVDNVD · Deferred4.0Critical9.1CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
EPSS (exploitation in 30 days)
0.39% · top 66% of all CVEs (16 Nov 2026)
CISA SSVC
Exploitation: none · Automatable: no · Technical impact: total
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: Critical 9.1.

Affected products

What to do

Solution: Baxter has released a software update for all impacted devices and software to address this vulnerability. A new version of the product that mitigates the vulnerability is available as follows: * Welch Allyn Connex Spot Monitor: Version 1.52.01 (available October 16, 2023) Baxter recommends users upgrade to the latest versions of their products. Information on how to update products to their new versions can be found on the Baxter disclosure page https://www.baxter.com/product-security  or the Hillrom disclosure page https://www.hillrom.com/en/responsible-disclosures/ . Baxter recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured and applied to the product (as described in the Connex Spot Monitor Service Manual).