CVE-2024-48973Debug port on Life2000 Ventilator serial interface is enabled by default
- 9.3CriticalCVSS base score
- 0.22%Exploitation in 30 daysEPSS · top 87% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Timeline
ID reserved
By Baxter
- 35 days later
· 35 days later
Published
- 4 days later
· 4 days later
Record updated
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| BaxterCNA (assigner) | 3.1 | Critical9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| NVDNVD · Deferred | 3.1 | Critical9.3 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.22% · top 87% of all CVEs (13 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: total
The list uses the assigner's score when there is one, then CISA's, then NVD's: Critical 9.3.
Affected products
Baxter · Life2000 Ventilation System
- affected 06.08.00.00 and prior
References
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2024-43157 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket