CVE-2025-14095Privilege boundary violation in Radiometer Products
- 6.8MediumCVSS base score
- 0.18%Exploitation in 30 daysEPSS · top 92% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
Timeline
ID reserved
By Radiometer
- 12 days later
· 12 days later
Published
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| RadiometerCNA (assigner) | 3.1 | Medium6.8 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| RadiometerCNA (assigner) | 3.1 | Medium5.7 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| NVDNVD · Deferred | 3.1 | Medium6.8 | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.18% · top 92% of all CVEs (18 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: partial
The list uses the assigner's score when there is one, then CISA's, then NVD's: Medium 6.8.
Affected products
Radiometer Medical Aps · ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- affected All application software versions with Windows 7, Windows XP as underlying OS
- affected Application software versions < 3.5MR11 with Windows 10 as underlying OS
Radiometer Medical Aps · AQT90 FLEX Analyzers
- affected All Application software versions <= 8.13 MR2
Radiometer Medical Aps · ABL800 BASIC and ABL800 FLEX Analyzers
- affected Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS
- affected Application software versions < 6.20 MR2 with Windows 10 as underlying OS
Radiometer Medical Aps · ABL9 Analyzers
- affected Application software versions < 1.5.0
What to do
Solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Workaround: Only authorized people can physically access the analyzer.
Exploitation: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.
References
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2025-203887 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://cveledger.app/cve/CVE-2025-14095)Record last updated .