CVE-2025-14096Credential Disclosure vulnerability in Radiometer Products

Description

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Timeline

  1. ID reserved

    By Radiometer

  2. · 12 days later

    Published

Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.

Scores

CVSS scores of CVE-2025-14096, by who scored it
Scored byCVSSScoreVector
RadiometerCNA (assigner)3.1High8.4CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NVDNVD · Deferred3.1High8.4CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS (exploitation in 30 days)
0.14% · top 96% of all CVEs (18 Nov 2026)
CISA SSVC
Exploitation: none · Automatable: no · Technical impact: total
Loading the EPSS history…

The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.4.

Affected products

What to do

Solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Workaround: Only authorized people can physically access the analyzer.

Exploitation: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.