CVE-2025-14096Credential Disclosure vulnerability in Radiometer Products
- 8.4HighCVSS base score
- 0.14%Exploitation in 30 daysEPSS · top 96% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Timeline
ID reserved
By Radiometer
- 12 days later
· 12 days later
Published
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| RadiometerCNA (assigner) | 3.1 | High8.4 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVDNVD · Deferred | 3.1 | High8.4 | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.14% · top 96% of all CVEs (18 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: total
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 8.4.
Affected products
Radiometer Medical Aps · ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- affected Windows 7 Operating system
- affected Windows XP Operating system
Radiometer Medical Aps · AQT90 FLEX Analyzers
- affected Windows 7 Operating system
- affected Windows XP Operating system
Radiometer Medical Aps · ABL800 BASIC and ABL800 FLEX analyzers
- affected Windows 7 Operating system
- affected Windows XP Operating system
What to do
Solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Workaround: Only authorized people can physically access the analyzer.
Exploitation: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
References
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2025-203889 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://cveledger.app/cve/CVE-2025-14096)Record last updated .