CVE-2025-14097Remote Code Execution Vulnerability in Radiometer Products
- 7.2HighCVSS base score
- 0.50%Exploitation in 30 daysEPSS · top 57% of all CVEs
- NoKnown exploitednot in CISA's KEV catalog
- NonePublic exploitsnone known yet
Description
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Timeline
ID reserved
By Radiometer
- 12 days later
· 12 days later
Published
Changes after publication (a new score, a public exploit, EPSS crossing 1, 10 or 50 %, leaving KEV) are kept since November 2026; earlier ones are not known.
Scores
| Scored by | CVSS | Score | Vector |
|---|---|---|---|
| RadiometerCNA (assigner) | 3.1 | High7.2 | CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H |
| NVDNVD · Deferred | 3.1 | High7.2 | CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H |
- EPSS (exploitation in 30 days)
- 0.50% · top 57% of all CVEs (18 Nov 2026)
- CISA SSVC
- Exploitation: none · Automatable: no · Technical impact: total
The list uses the assigner's score when there is one, then CISA's, then NVD's: High 7.2.
Affected products
Radiometer Medical Aps · ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- affected Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems
- unaffected Application software versions >= 3.5MR11 with Windows 10 operating system
Radiometer Medical Aps · ABL800 BASIC and ABL800 FLEX Analyzers
- affected Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems
- unaffected Application software versions >= 6.20MR2 with Windows 10 operating system
Radiometer Medical Aps · AQT90 FLEX Analyzers
- affected Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems
- unaffected Application software versions >= 8.13 MR2 with Windows 10 operating system
What to do
Solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Workaround: If the network is not considered secure, please remove the analyzer from the network.
Exploitation: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
References
Sources
The same CVE in the primary databases and aggregators, to check against.
- CVE.org record (opens in a new tab)
- CVE List JSON (GitHub) (opens in a new tab)
- NVD (opens in a new tab)
- FIRST EPSS (opens in a new tab)
- EUVD EUVD-2025-203895 (opens in a new tab)
- OSV.dev (opens in a new tab)
- GitHub Advisories (opens in a new tab)
- CIRCL Vulnerability-Lookup (opens in a new tab)
Badge for a README or a ticket
[](https://cveledger.app/cve/CVE-2025-14097)Record last updated .